Mmicroscope = Mobj х Madd х Meyepiece, where
Mobj – for a flat-field microscope, this is objective magnification, x (indicated on the objective).
Mobj – for a stereomicroscope with the Greenough optical design: product of the zoom objective magnification (x, indicated on the magnification change knob) and magnification of an auxiliary objective lens (x, indicated on the lens).
Mobj – for a stereomicroscope with the CMO design: product of the zoom objective magnification (x, indicated on the magnification change knob) and magnification of the common main objective (x, indicated on the common main objective).
Madd – magnification of additional optics:
magnification of the microscope head (specified in the microscope certificate), or
magnification of an additional intermediate attachment (specified in the certificate and on the attachment), or
magnification of an additional built-in lens (specified in the microscope certificate and on the microscope).
For instance, the MAGUS Metal V790 DIC microscope allows the introduction of a 1.5x magnification lens into the optical path.
Meyepiece – eyepiece magnification (x, indicated on the eyepiece).
Resolution limit is the shortest distance between two points (lines) that can still be distinguished and do not merge into one. The points spaced closer than the microscope’s resolution limit cannot be distinguished and will appear as a single point.
The resolution limit is calculated by the formula d = λ/2NA, where λ is the light wavelength, and NA is the numerical aperture of the objective. The wavelength of white light is 0.55µm.
Working distance is the distance between the specimen surface or the cover slip surface (if there is one) and front edge of the objective.
Depending on the objective design, it may be a front lens or a metal frame of the objective. The cover slip thickness is engraved on the objective.
Modern microscopes are most commonly designed for objectives with a parfocal distance of 45mm. However, superior aberration correction requires more complex optics, resulting in a longer objective and a greater parfocal distance.
For instance, modern plan apochromatic objectives frequently feature a 60mm parfocal distance.
The parfocal distance is defined when the image is in sharp focus.
>All of the objectives in a given microscope have the same parfocal distance. Therefore, only fine focusing adjustment is required when the revolving nosepiece is rotated.
The proper set-up of Köhler illumination offers the following benefits:
– the highest possible resolution on each objective
– focusing on the specimen image, removing the images of artifacts: dust on the illuminator or on the slide, glare
– even illumination of the entire field of view with no edge darkening.
The Zernike condenser is a phase contrast condenser named after Frits Zernike, a Dutch physicist.
It is designed to visualize transparent unstained samples (cells and microorganisms). The phase contrast microscopy converts invisible phase shifts into visible variations of brightness.
The condenser is compatible with the MAGUS Bio 250, 260, 270, and 290 biological upright microscopes as well as the MAGUS Lum 400, 400L, and 450 fluorescence upright microscopes.
The Zernicke condenser comes standard with the MAGUS V350 inverted biological and the MAGUS Lum V500/MAGUS Lum V500L inverted fluorescence microscopes.
Epi-fluorescence illuminators are employed as reflected light sources in biological microscopes.
Transparent unstained specimens do not absorb light, resulting in low image intensity contrast. Brightfield observations do not allow you to view transparent specimens clearly. Phase contrast microscopy is designed to visualize transparent, unstained samples.
When light passes through transparent specimens, there are minor phase shifts that a human eye cannot see. Phase contrast microscopy converts these phase shifts into amplitude changes. We can view these amplitude changes as the different contrast of the image.
The phase contrast device is included in the standard kit of inverted biological microscopes, as these models are most commonly used to examine live specimens in their natural state.
Biological upright microscopes do not have the device in a standard delivery, as these microscopes are primarily designed for the brightfield observation of stained specimens.
– a Zernike phase condenser (replacing the Abbe condenser) or a slider with phase rings (inserted into the slot of the standard Abbe condenser)
– a set of phase objectives
– a centering telescope (replacing the eyepiece to adjust the condenser phase ring with the phase objective ring).
These accessories are typically combined into a single phase contrast device and sold as a complete set.
The differential interference contrast (DIC) is a microscopy technique that converts light phase shifts into visible variations of brightness and contrast. The technology is based on splitting a polarized beam into two coherent beams that pass through adjacent areas of the specimen. The difference in the optical density of the specimen creates a phase shift between the beams, which, once they are recombined, is converted into a clear three-dimensional image with a typical pseudo-relief effect.
The DIC microscopy is indispensable for studying transparent unstained structures, such as living cells, thin films, or crystals. Unlike phase contrast, it does not create halo artifacts, thereby ensuring the clarity of edges due to the use of polarized light. The technique finds applications in biology, materials science, and medical diagnostics, enabling the study of microstructures without disrupting their natural state.
Biological inverted microscopes designed to study samples in special glassware with a bottom thickness of up to 1.2mm come pre-equipped with a phase contrast device.
We recommend the following configuration:
– the MAGUS Bio 250TL microscope
– the MAGUS PH1 phase contrast device
– the MAGUS CHD40 digital camera
While phase contrast microscopy provides superior diagnostic information, phase contrast microscopes involve significantly higher costs compared to darkfield ones.
We recommend the following options:
1. The MAGUS Bio D230TL LCD + oil darkfield condenser + 0.5х or 0.65х C-mount adapter.
2. The MAGUS Bio 250TL + oil darkfield condenser + the MAGUS CHD40 camera + the MAGUS MCD40 monitor.
3. The MAGUS Bio 270T + phase-contrast device + the MAGUS CHD40 camera + the MAGUS MCD40 monitor.
First-choice microscopes: MAGUS Bio 230 and 240 series.
Second-choice instrument: MAGUS Bio 250 and 260 series. Third-choice option: MAGUS Bio 270 model.
We can also recommend complete solutions for specific diagnostic applications:
General histology, mites, and skin scrapings:
– the MAGUS Bio 250B + the MAGUS DFS1 Darkfield Slider + other accessories as necessary
– the MAGUS Bio 250BL + the MAGUS DFS1 Darkfield Slider + other accessories as necessary
Bacteriological laboratory:
– the MAGUS Lum 400L + the MAGUS PH1 Phase Contrast Device
Leptospirosis: darkfield:
– the MAGUS Bio 250B + the MAGUS DFS1 Darkfield Slider
– the MAGUS Bio 250BL + the MAGUS DFS1 Darkfield Slider
Rabies:
– the MAGUS Lum 400L + the MAGUS PH1 Phase Contrast Device
Lichen:
– the MAGUS Lum 400
– the MAGUS Lum 400L
The MAGUS Stereo 7T is up to the task.
It can be upgraded with the optional MAGUS CHD10 camera and MAGUS MCD20 monitor, requiring a 0.5x C-mount adapter (MAGUS ZFA050) for installation.
We also recommend that you check out our Complete Solutions section.
The MAGUS Bio 230B:
– The Köhler illumination guarantees the highest resolution on each objective and the required contrast of the specimen image. Achromatic objectives correct field curvature by 65%.
– Eyepiece tubes are 360° rotatable, which allows for adjusting the height of the eyepieces.
– The revolving nosepiece is oriented toward the interior, which frees up more space above the stage. The user can see the objective introduced into the optical path.
– The fifth free slot of the revolving nosepiece is intended for an additional objective.
– The stage has no X-axis rack and pinion. The belt-driven mechanism allows for smooth movement and improves the comfort of use: You will not accidentally scratch your hand on the rack, for example.
– The Abbe condenser is designed to install a darkfield or phase-contrast slider, which saves time when switching microscopy techniques.
– A variety of accessories expands the choice of microscopy techniques and the microscope magnification range.
The MAGUS Bio 270T:
– In the conventional Köhler illumination design, the light source is brought out from underneath the collector; in this model, it is moved to the rear of the base. The design ensures even illumination, the highest resolution on each objective and the required contrast of the specimen image. Plan achromatic objectives correct field curvature by 90% when eyepieces with a 22mm field of view are used.
– Eyepiece tubes are 360° rotatable, which allows for adjusting the height of the eyepieces.
– The user programs the revolving nosepiece during setup, which improves comfort and saves time when switching objectives: The microscope remembers the brightness selected for each objective and automatically adjusts the value when the revolving nosepiece is rotated.
– The revolving nosepiece is oriented toward the interior, which frees up more space above the stage. The user can see the objective introduced into the optical path.
– The fifth free slot of the revolving nosepiece is intended for an additional objective.
– The stage has no X-axis rack and pinion. The belt-driven mechanism allows for smooth movement and improves comfort of use: You will not accidentally scratch your hand on the rack.
– The microscope’s intelligent mechanism controls the illumination: It automatically selects the brightness when objectives are changed, switches off the microscope by a timer, and displays the operation status on the LCD screen. The user can adjust the desired color temperature and lock the brightness adjustment, as necessary.
– The hidden placement of the power adapter and power cord enhances the aesthetics of the workstation and simplifies its storage. In combination with a special carrying handle, the microscope can be carried more easily and securely.
– A variety of accessories expands the choice of microscopy techniques and the microscope magnification range.
It includes the following objectives:
4x NA=0.10, WD=30mm
PH 10x NA=0.25, WD=10.2mm
PH 20x NA=0.40, WD=12mm
PH 40x NA=0.6, WD=2.2mm
WD is the working distance.
We suggest considering the following package:
– MAGUS Bio 250T or 250TL Biological Microscope
– MAGUS PH1 Phase Contrast Device
– MAGUS CBF50 Digital Camera + MAGUS CFA050 0.5x C-mount adapter or
MAGUS CBF90 Digital Camera + MAGUS CFA065 0.65х C-mount adapter
another option for brightfield observations: MAGUS Bio 270T + MAGUS CHD series digital camera.
The MAGUS V500 and V500L models are suitable for this.
Its main features are:
– Microscope head with the trinocular tube for mounting a digital camera and displaying an image on the monitor screen. The inclination angle is adjusted to match the user’s height. Three modes of beam splitting: 0/100, 100/0, or 80/20.
– Strain-free plan semi-apochromatic and plan apochromatic objectives introduce no optical artifacts, thereby ensuring the high resolving power of the microscope.
– Transmitted and reflected light for work with transparent and opaque samples.
– Powerful 100W halogen illuminators of transmitted and reflected light for bright natural lighting.
– Orthoscopic examinations, Bertrand lens for conoscopic examinations, compensators, Köhler illumination in transmitted and reflected light.
– Convenient alignment of the revolving nosepiece slots and stage.
– Rotatable round stage with a vernier scale for accurate measurements of the rotation angle can be locked in the desired position.
Stereomicroscopes are used for examining whole mineral specimens. We recommend the MAGUS Stereo 7 and Stereo 8 series with a darkfield condenser and a polarizer/analyzer set.
Polarizing microscopes are used for studying thin sections and polished sections in petrography, geology, and mineralogy.
Check out the section with the polarizing microscopes of MAGUS Pol series.
For pearlescent paint analysis, we recommend using not only brightfield microscopy but also the polarization technique: The MAGUS Metal 600 microscope is well suited for this application.
If you need to study scratches on coatings, darkfield microscopy may be particularly useful. In this case, we recommend using the MAGUS Metal 600 BD model that employs three techniques: brightfield, darkfield, and polarization microscopy.
1000x magnification is sufficient. Phase contrast microscopy or DIC (differential-interference contrast) technique will reveal the features of ornamentation and other details.
The phase contrast device is compatible with the MAGUS Bio 250, 260, 270, and 290 biological upright microscopes.
The DIC device is compatible with the MAGUS Bio 290.
A stereomicroscope, for example, the MAGUS Stereo 7B, is suitable for analyzing the macrostructure.
A metallurgical microscope is well-suited for examining the microstructure of metals and alloys. The microscopy techniques – brightfield, polarization, darkfield, DIC – are selected based on the samples and tasks to be solved. All of the MAGUS metallurgical microscopes are given in this section.
For example, nylon fiber:
– Fig. 1 – Observation in plane polarized light reveals differences in the refractive indices of the fiber and the presence of opaque particles.
– Fig. 2 – Cross-polarized light reveals third-order polarization colors, whose distribution within the fibers indicates that this is a cylindrical rather than irregular fiber, which can be used to determine mechanical strength.
– Fig. 3 – The introduction of a quartz wedge into the optical path enables the determination of optical path differences for measuring birefringence.
If no in-depth fiber analysis is expected, you can use the MAGUS Metal 600 metallurgical microscope to examine the sample surface. The microscope features the polarization technique.
We recommend the MAGUS Metal D630 LCD that is equipped with a digital camera and an LCD monitor. The original design of the stand and focusing mechanism does not limit the height of the sample.
The examined surface of the sample should be:
– parallel to the base surface (stage plane)
– perfectly aligned (not tilted or uneven)
A stereomicroscope should be used to examine wood for defects. If the wood piece is large, a stereomicroscope on a universal stand works well, e.g., one of our complete solutions.
The microscopy technique is brightfield.
Anatomical analysis of wood using optical microscopy involves preparing very thin, transparent sections and observing them in transmitted light with an optical microscope. This technique enables the examination of wood structure and identification of the species composition based on diagnostic features. Observations are performed at magnifications of 50–120x, while fiber type identification based on morphological features requires magnifications of 200–500x.
The MAGUS Metal 600 is suitable for such applications, as it operates in both transmitted and reflected light.
You should also purchase an optional 80x MAGUS 80PLL objective for it. The resolution limit of an 80x/0.8 objective is 0.35μm, while the depth of focus is 0.4μm. This magnification is enough to see 1 micron microspheres. For 200 micron microspheres, use the 5х or 10х objective.
The MAGUS CHD40 digital camera features an 11.14x6.26mm sensor. Therefore, the field of view on this microscope when the image is displayed on a monitor is: 15.9x8.9mm with a 0.7x objective magnification and 1.7x0.99mm with a 6.3x objective magnification.
The laws of optics do not allow both high magnification and a large field of view at the same time; therefore, it is impossible to make measurements with an accuracy of 0.01mm with a large field of view.
For this type of measurements, use the MAGUS CSS1 calibration slide for stereomicroscopes.
We recommend the following accessories:
– the MAGUS 100PLL objective if there is no 100x objective in the standard delivery – to achieve the highest resolution
– 10x eyepiece with reticle – for measurements
– 10x eyepiece with a scale – for analyzing the metal grain size per the ASTM E112 standard
– calibration slide – to calibrate the eyepiece scale for each objective
For an inverted metallurgical microscope:
– the MAGUS MD20 eyepieces – with diopter adjustment
The visualization system can be configured with the MAGUS CHD20 digital camera and MAGUS MCD20 monitor.
For research, you should consider a set of the Levenhuk N80 NG prepared slides “See it all”, Levenhuk G100 cover slips, and Levenhuk G50 blank slides.
If you apply the resolution limit formula to a 4x/0.1 objective where:
d is the linear resolution limit of the microscope
λ is the wavelength of the light that illuminates the object
NA is the numerical aperture of the objective
You will get the resolving power of 2.75µm. Then, convert it to millimeters and get 0.00275mm.
The useful magnification of a microscope when a 4x/0.1 objective is used is approximately 50–100x. This magnification is achieved with 12.5х–25х eyepieces.
You should choose optional eyepieces from this range with the maximum field of view for the user’s convenience.
The MAGUS Bio series includes the following models: Bio 230, Bio 240, Bio 250, Bio 260, Bio 270, Bio 290.
The MAGUS Bio 250 kit includes plan achromatic objectives and allows for setting up Köhler illumination that guarantees the highest resolution on each objective and the required contrast of the specimen image. While there is no intelligent lighting control system in the microscope, optional accessories will enable not only brightfield observations, but also darkfield, phase contrast, and polarization techniques.
If needed, you can purchase this eyepiece separately. When 40x/0.65 and 100x/1.25 (oil) objectives are used, magnification of the 25x eyepiece is “useless” as it provides no new detail in the image of the sample.
Objectives of an inverted microscope have an increased focal length. The optical design of a 100x objective will not allow it to focus on the sample due to the thickness of the dish bottom.
The lighting system of an inverted biological microscope will not allow you to apply light at an angle required for a 100x objective.
MAGUS Bio 260T:
– the MAGUS CHD20 camera + the MAGUS MCD20 monitor
– the MAGUS CHD40 camera + the MAGUS MCD40 monitor
MAGUS Bio 270T:
– the MAGUS CHD40 camera + the MAGUS MCD40 monitor
– the MAGUS CHD50 camera + the MAGUS MCD40 monitor
1. Colonies suspected of being Legionella are detected when viewing dishes on a stereomicroscope, and so we can recommend the MAGUS Stereo 9T model.
2. Gram staining is used to identify the bacteria of Legionella spp. MAGUS brightfield biological microscopes are suitable for such research.
3. The identification of Legionella bacteria using the fluorescent antibody technique is carried out on a fluorescence microscope in blue light. A specimen is prepared on the slide and stained with a fluorochrome.
The tasks from items 2 and 3 are covered by 3 microscope models: MAGUS Lum 400, Lum 400L, and Lum 450L.
The MAGUS Lum 400 is equipped with a 100W mercury lamp.
The reflected light source of the MAGUS Lum 400L is four 3W LEDs of various wavelengths.
For educational purposes, we recommend the MAGUS CLM10 camera.
The MAGUS CLM30, CLM50, CLM70, and CLM90 models, featuring a Peltier cooling sensor, are designed for professionals to substantially reduce image noise. To solve the tasks with low-light conditions, we recommend using a monochrome camera.
The Common Main Objective (CMO) design, also known as Abbe or Galilean optics.
The objective is exactly perpendicular to the specimen plane.
The objective diameter limits the stereoscopic angle to 11 degrees.
There are two independent parallel optical channels above the objective and they are also perpendicular to the specimen plane. The system is designed to project the specimen image to infinity.
Benefits of the CMO design:
− Large field of view
− Since the optical axis of the objective is perpendicular to the specimen plane, there are no image distortions caused by the tilted optical axis in Greenough systems
− High-end stereomicroscopes utilize plan apochromatic (Plan-Apo) objectives. Chromatic aberrations and field curvature are fully corrected. This results in high image quality both in the eyepieces and on the camera.
Check out the MAGUS Stereo 8T.
You will need to also purchase two 30x eyepieces (MAGUS SD30 30х/8mm) and a 2x auxiliary objective lens (MAGUS AL20), which will allow you to achieve the maximum magnification of 330x.
As the splitting is permanent, there is no need to manually switch the prism, which helps avoid damage to the microscope head. The light beam to the camera port comes from the right channel of the Greenough optical system. A total of 80% to the right eyepiece and 20% to the camera port.
The user can observe an image in the eyepieces and on the screen at the same time. The difference in light intensity in the left and right channels is negligible, ensuring user comfort. A total of 20% of light in a camera port is sufficient for MAGUS light-sensitive cameras.
Adapters are installed on these attachments:
– MAGUS CMA050 A10 C-mount adapter
– MAGUS CMA100 A10 C-mount adapter
The camera is threaded onto the adapter.
If not, you can install the camera instead of the eyepiece using eyepiece adapters and connectors. To choose an adapter, you need to know the eyepiece tube diameter and the size of the selected sensor.
For more information, check out the camera selection guide.
Reduction will depend on the camera sensor size is – the larger the sensor, the greater the reduction will be.
Approximate values:
Camera sensor size Recommended adapter
1/4"" 0.265x or 0.3x
1/3"" 0.3x
1/2"" 0.45x, 0.5x, or 0.6x
1/1.8"" 0.45x, 0.5x, or 0.6x
2/3"" 0.7x
MAGUS supplies professional equipment, which does not include training cameras with small 1/4"" and 1/3"" sensors. Therefore, there are no adapters with a magnification factor of less than 0.37x in our range.
Diagonals should be counted in millimeters. You cannot use inches in calculations because monitor inches and camera vidicon inches are not equivalent units.
1. The MAGUS CHD40 digital camera – it does not require an additional adapter because it has a large sensor.
2. The MAGUS CHD50 digital camera + MAGUS CFA065 0.65x C-mount adapter.
Requirements to the panel: HDMI port and 4K resolution.
Always verify the eyepiece diameter. It measures 23.2mm for the MAGUS Bio 240B model.
For camera installation, you’ll need one of the following C-mount adapters: MAGUS CMT037, CMT050, CMT075, or CMT100.
The adapter magnification is selected to match the camera sensor size and application requirements – whether to maximize the field of view on screen or achieve maximum magnification.
1. MAGUS CBF70 digital camera + MAGUS ZFA100 1x C-mount adapter.
2. MAGUS CDF70 digital camera + MAGUS ZFA100 1x C-mount adapter.
A budget-friendly option with high resolution and a compact sensor size:
1. MAGUS CBF10 digital camera + MAGUS ZFA050 0.5x C-mount adapter.
We suggest considering a more advanced option – MAGUS CBF30 digital camera:
1. The USB 3.0 port provides high-speed data transmission.
2. A larger sensor has a number of benefits:
– less noise, resulting in cleaner images
– more light capture, delivering richer tones and greater color depth
– larger field of view coverage
– the larger the sensor is, the more that larger photodiodes will fit, which means that the two tasks of increasing the light sensitivity and resolution are solved.
Since the resolution limit of the 40x objective is about 0.42µm, we recommend an additional 16x eyepiece, which provides the upper limit of useful magnification on this objective.
A standard 10x eyepiece is sufficient for a 100x objective.
If you need polarization and phase contrast, you should choose a biological microscope with optional accessories that will allow for both microscopy techniques.
All manufacturers have unique “objective–condenser (slider)” sets. We can only suggest MAGUS-brand microscopes equipped with phase contrast devices.
A microscope is an optical instrument. The microscope objective produces a magnified image of the specimen. Then, this image must be directed somewhere. It is routed either to the eyepieces or to a digital camera.
The eyepieces further magnify the image, allowing us to observe it visually. To view an image on the screen, it needs to be digitized. The camera captures the image that is produced by the objective, converts it into digital format (photo/video), and then transmits it to the monitor.