Description
DIC condenser
The DIC condenser is installed in place of the brightfield condenser. The turret includes DIC prisms and a slot that is used during brightfield observations. The DIC prisms in the turret are used depending on the objective being used.
DIC sliders
The DIC slider is installed in a special slot above the revolving nosepiece. The DIC slider is used in combination with the corresponding objective. The slider is a plate with a DIC prism and an analyzer. The MAGUS DIC290 device includes two sliders: One is used with 10x and 20x objectives, and the other one with 40x and 100x objectives. The rotation of the slider knob moves the DIC prism in the transverse direction.
Polarization filter
When the DIC technique is used, the light from the illuminator is passed through a polarization filter. The polarizer is placed on the collector.
Recommended objectives
The objectives are not included in the kit of the MAGUS DIC290 device.
The DIC technique requires a precise position of the optical components relative to each other: the lower DIC prism – relative to the front focal plane of the condenser and the upper DIC prism – relative to the rear focal plane of the objective. The MAGUS DIC290 device is designed for MAGUS Bio 290T microscope and, therefore, you should use the microscope’s supplied or compatible objectives for DIC observations.
We recommend using MAGUS FL S-APO60 PlanF objectives with a higher aperture. A higher aperture of an objective allows focusing on a thin flat section of a thick specimen, which is called optical sectioning. Higher apertures also increase the resolving power of the optical microscope.